GE RAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
5-60 Section 5-15 - Boot Screen Functions
Section 5-15
Boot Screen Functions
5-15-1 Overview
NOTE: Prior to the “Boot Screen” ( see figure below), the BIOS version is shown on the monitor:
Following LINUX supported functions are available as soon as the “Boot screen” appears:
5-15-1-1 Voluson
The System will boot-up in windows. The Ultrasound Application is started.
For details refer to Section 3-6-2-3 "During a normal boot, you may observe" on page 3-27.
5-15-1-2 GE-Service
The Voluson Maintenance System functions are secured by service password (class C).
1.) Turn system OFF and then back ON.
2.) As soon as the “Boot Screen” appears (see: Figure 5-33 above), press the PG DN [ Arrow
down] key on the keyboard until the GE Service item is highlighted, then press ENTER.
3.) Enter the Password Rudi (case sensitive) and confirm with ENTER. The system boots into
LINUX.
Figure 5-33 Boot screen
!! NOTICE:
After 3 sec. without pressing any key, the system will boot-up in windows (= Voluson item). If you
missed selection, retry again with Ctrl+Alt+Del.
After 3 sec. without pressing any key, the system will boot-up in windows (= Voluson item).
If you missed step 2, retry again with CTRL + ALT + DEL.
Voluson
GE-Service
Rolback
Memtest(Memory Check)