Chapter 5 - Components and Functions (Theory) 5-15
5-2-3-5 CFM-Mode (Color Flow Mode)
1.) DRFG
- PRF-generator; see: Section 5-2-3-1 "B-Mode" on page 5-12
- After DC-cancelling the signal is mixed with RX-Frequency and brought to LF-Spectrum,
where the LowPassFilter cuts HF.
Mixer and Magnitude-Calculator arrange Complex Demodulation.
5-2-3-6 3D-Mode (Freezes after 1 volume sweep)
see: Section 5-2-3-1 "B-Mode" on page 5-12
5-2-3-7 Real Time 4D-Mode (nonstop volume rendering)
see: Section 5-2-3-1 "B-Mode" on page 5-12
5-2-3-8 XBeam CRI-Mode (CrossBeam Compound Resolution Imaging)
see: Section 5-2-3-1 "B-Mode" on page 5-12
5-2-3-9 Archive write mode (store Image to Archive)
1.) DRFG
Not used