GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-115
Section 8-36
Replacement of the Caster Cap BT16
8-36-1 Manpower
One person, 2 minutes
8-36-2 Tools
None
8-36-3 Preparations
1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
NOTE: Lock Caster before removing it.
8-36-4 Removal Procedure
1.) Remove the FRONT CABINET BEZEL ASSY BT16. Refer to the figure below.
Figure 8-168 Removing the FRONT CABINET BEZEL ASSY BT16
FRONT CABINET BEZEL
ASSY BT16