GE RAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
7-40 Section 7-6 - Common Service Desktop (CSD)
7-6-6-2 Scanner Utilities
Device Listener Service
This will analyze the USB connect/disconnect logs from the Device Listener Server. The output is
displayed on the screen.
Device Listener Service Log
Shows USB "disconnects".
Start Rollback
Operation see Section 5.17.1.3 "Rollback" on page 5-64
Start Restore
Operation see Section 5.17.5.2 "Restore" on page 5-72
System Shutdown
System Shutdown is intended for use from a remote computer. The menu gives you the ability to either
Restart or Shutdown the system.
□ Force FastPoll after restart
- Should be checked if you are working from a remote computer.
- Should be unchecked if you are working locally on the Voluson S&P-Series system.
□ Keep Disruptive Mode after restart
- Should be checked if you are working from a remote computer.
- Should be unchecked if you are working locally on the Voluson S&P-Series system.
NOTE: After performing any diagnostics, always disable "Disruptive Mode" and reboot the system before
returning it to customer use. Operation see Section 7.6.9 "After performing Diagnostics" on page 7-48 .
NOTE: When accessing through InSite (remotely), a system shutdown or restart is prohibited, whenever
Disruptive Mode status is disabled.
Dicom Verify
This enables to verify DICOM devices.