EasyManuals Logo

GE Voluson P6 BT18 User Manual

GE Voluson P6 BT18
620 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #218 background imageLoading...
Page #218 background image
GE RAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
5-14 Section 5-2 - General Information
5-2-3-4 D-Mode Autotrace (draws PC-calculated envelope to D-Spectrum)
1.) DRFG
- PRF-generator; see: Section 5-2-3-1 "B-Mode" on page 5-12
- After DC-cancelling the signal is mixed with RX-Frequency and brought to LF-Spectrum,
where the LowPassFilter cuts HF.
Mixer and Magnitude-Calculator arrange Complex Demodulation.
a.) DMA section
D-mode-Data is written via SP0 into SDRAM Fifo Buffer memory.
PC calculates Autotrace-Curve from D-Mode data.
Cine Mode with Autotrace: Cine Mode-Memory for the Autotrace-Curve is inside SOM-
Memory.
Software has to take care that D-Spectrum and Autotrace-Curve are placed exactly one upon
the other, means: have the same Cine-Shift.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson P6 BT18 and is the answer not in the manual?

GE Voluson P6 BT18 Specifications

General IconGeneral
BrandGE
ModelVoluson P6 BT18
CategoryMedical Equipment
LanguageEnglish

Related product manuals