GE RAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
8-132 Section 8-42 - Replacement of the DRFG
8-42-6 Serial Number - Reprogamming Procedure
Before calling OLC please prepare the following data:
• SNO of Voluson™ P8/Voluson™P6 (Alphanumeric serial number can be found on main label at
back of system)
• Hardware ID number (See SYSTEM INFO page or label of DRFG board)
1.) Gather System Serial number and Hardware ID
- Press the F2
key on upper OPIO, select SYSTEM SETUP, select ADMINISTRATION, click
on SYSTEM INFO and write down the "Hardware ID" number.
!! NOTICE:
To reprogramming the System Serial Number, contact the OLC(On Line Center) to get the necessary
"Serial Key" and "Permanent Key" (calculated in OKOS) !!!
Figure 8-191 System Setup - Administrator - System Info page
!! NOTICE:
The Hardware ID can also be found on the label on the DRFG board.