GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-171
Section 8-56
Replacement of CWD piggy
8-56-1 Manpower
One person, 30 minutes.
8-56-2 Tools
Philips screwdriver, Hirose ejector
8-56-3 Preparations
Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
8-56-4 Removal Procedure
1.) Unscrew 2 screws and remove POWER CORD FIX BRKT and power cord.
Figure 8-244 Power Bracket
POWER CORD FIX BRACKET
power cord