4 Setup
Operator’s Manual 4 - 39
4.6.8 Q-Path Preset
Q-path is a network server provided by Telexy Healthcare Inc. for digital image storage. Q-View is
a client viewing tool for the server. Telexy Healthcare developed technology and a command
structure that allows any Q-view enabled ultrasound system to access Q-path directly from the
ultrasound system using a single control. The primary purpose for Q-view is to provide remote
access to Q-path from the ultrasound system to complete the exam report on the ultrasound system
and submit for QA,
For details, please contact Q-Path service provider.
Perform the following procedure:
1. Select “Enable Q-Path”.
2. Enter the website, account and password of the target service.
3. Select user type: Personal User or Default User.
– Personal User: the personal user needs to enter the user name and password in every-time
login.
U-VIEW Server
Setup
Device Tap to input the ID of device.
IP Address Tap to input IP address of U-VIEW server.
Port Tap to input port of U-VIEW server.
Connect Select to connect the server.
Type Item Description
Item Description
Advanced Sets the sub URLs of “QView full” and “QView lite”.
The sub URL is set by default. Users can modify the sub URL and click
[Apply] to exit the “QView sub URL setting” window.
Worksheet Only Sets whether to directly enter the Worksheet interface after opening the
Q-Path server.
Password On Worksheet Sets whether to display the Signature field box in a worksheet.
Tap [Report] > [WorkSheet] or tap [Review] > [Report] > [WorkSheet],
enter the worksheet password in the field box, and click [OK].
Users can query the corresponding worksheet by searching the
worksheet password in the Q-Path server.
Password On End Exam Sets whether to input the worksheet password after ending an exam.
Password Visible Sets whether the password is visible.
Import Imports a user-defined worksheet template from the USB storage
(downloaded from the Q-Path server).
Backup Backs up worksheets to the USB storage.
Restore Restores the backup worksheet template from the USB storage to the
ultrasound system.
Delete Deletes a worksheet template.
Restore Factory Restores the worksheet template to the default state.