13 - 56 Operator’s Manual
13 Probes and Biopsy
13.2.7 Disposal
Be sure to sterilize the needle-guided bracket before disposing of it.
Disposal of the needle-guided brackets is in accordance with SanPiN 2.1.7.2790-10.
Contact your MINDRAY representative when disposing of this device.
13.3 Middle Line
Middle Line helps to locate and observe the focus point of lithotripsy wave during lithotripsy
treatment. By means of providing information for the lithotripsy machine as well as a tool for
watching the procedure of lithotripsy in real-time, you can adjust the intension and frequency of the
lithotripsy wave through lithotripsy machine.
NOTE:
This function in the ultrasound system is for lesion (stone) location and observation only. For
details, please refer to lithotripsy machine accompanying manuals.
Turn on Middle Line function.
The middle line is a vertical dotted line located in the middle of the screen, the position and
direction of which cannot be changed.
There is a mark icon of “×” located on the middle line which can be moved up and down along
the line.
To use the Middle Line function of the ultrasound system:
a. Change the mark position and by adjusting lithotripsy machine tools or patient posture to
locate the stone center at the mark.
b. Read the depth of the mark by observing the depth caliper on the screen.
c. After the stone is located, refer to lithotripsy machine manuals to perform the lithotripsy.
d. The depth of the mark is displayed in the image parameter area of the screen.
13.4 eSpacial Navi
The eSpacial Navi function builds up the connection between the ultrasound system and the
processed needles. The needle position appears on the image in real time. Meanwhile, the virtual
needle mark guides the needle path on the ultrasound image. Based on the magnetic induction
technology, the eSpacial Navi function aids in and enhance the ultrasound needle guidance.
WARNING
Ultrasound images should be referenced during the whole process of
needle guidance.
The magnetizer should be used with the needle cap. Disinfect the
magnetizer after use.
The magnetizer is a single use device. Dispose of the magnetizer properly
after use.
The needle cap is a sterile, single use device. To maintain procedural
sterility, it is important that the needle cap is isolated from non-sterile
objects.

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals