8 - 6 Operator’s Manual
8 Smart 3D
Scanning plane and probe movement
Move the probe across the body surface.
The arrow in the figure below indicates the movement of the probe.
You can move the probe in the opposite direction to the arrow.
8.3 Smart 3D Image Acquisition
NOTE:
In Smart 3D image scanning, if the probe orientation mark is oriented to the operator’s finger,
perform the scan from right to left in linear scan, or rotate the probe from left to right in rocked
scanning. Otherwise, the VR direction will be wrong.
Perform the following procedure:
1. Select the proper probe and exam mode; make sure there is sufficient gel on the probe for
scanning.
2. Obtain a 2D image, and optimize the image if necessary.
3. Tap [Smart 3D] to enter Smart 3D imaging mode.
– Adjust ROI size and position and the position of VOI.
– Select the render mode.
– Set the scan method and the movement of the probe.
4. Tap [Update] to start the 3D image acquisition.
The system enters into 3D image review status when the acquisition is completed; or, you can
finish the acquisition ahead by tapping [Freeze] or [Update].
5. Exit Smart 3D
– Tap [Update] or [Freeze] to return to Smart 3D acquisition preparation status.
–Or tap [B]
to exit the mode.

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals