EasyManuals Logo
Home>Mindray>Diagnostic Equipment>TEX20

Mindray TEX20 Operator's Manual

Mindray TEX20
330 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #251 background imageLoading...
Page #251 background image
13 Probes and Biopsy
Operator’s Manual 13 - 57
CAUTION
Keep the magnetizer away from any device that is sensitive to magnetic
fields. Never place or hold the magnetizer within 1 m of the probe.
The eSpacial Navi system supports only the needle types listed on the
needle list. For details, see “13.4.1 Interface”.
Keep the eSpacial Navi probe away from the source of disturbance, such
as metal objects and magnetized objects. Do not use or store the eSpacial
Navi probe in the vicinity of strong electric or magnetic fields.
NOTE:
When used under optimal operating conditions by an experienced medical professional, the
eSpacial Navi system can achieve precision as high as ±2.4 mm.
Place the magnetizer in the proper position to prevent the magnetizer from falling off or
becoming damaged.
DO NOT make the magnetizer overheated (more than 50 °C) during cleaning and
disinfections. High temperature may cause the magnetizer to become deformed or damaged.
13.4.1 Interface
During biopsy, the selected needle type and prompt icons are displayed on the screen. The meaning
of prompt icons are as follows:
Alignment indicator (plan view)
Alignment indicator (plan view)
Filed strength
Maximum depth at which the needle can be detected
Appears when the insertion angle of a needle relative to the skin
surface exceeds 60° to notify users to adjust the angle.

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals