6 - 10 Operator’s Manual
6 Image Acquisition
Packet Size
This function is an indication of the ability to detect flow, which is used to adjust the accuracy of
color flow.
The higher the sensitivity is, the more sensitive indication for low-velocity flow becomes.
Dynamic Range
This function is to adjust the transformation of echo intensity into color signal.
Increasing dynamic range will lead to higher sensitivity to low-power signals, thus enhances the
range of signals to display.
iTouch
To optimize image parameters as per the current tissue characteristics for a better image effect.
Smart Track
To optimize image parameters as per the current tissue characteristics for a better image effect.
When Smart Track is turned on, the system optimizes ROI angle and position automatically to
achieve an active tracking by reducing the impact of patient respiratory movement.
Dual Live
This function is used to display Power image synchronously.
Dual
To enter the Dual-split display mode, and to switch between the windows.
V 1:1
This function is to display images in vertical format in the dual-split mode. After the feature is
enabled, one image appears above, and the other image appears below.
In the dual-split mode, tap [V 1:1] to enable this function.
HR Flow
Enhance tiny vessel display to analyze the blood supply of the vessel in pathological organ.
Tap [HR Flow] to complete the adjustment ([HR Flow] is highlighted after it being enabled).
6.4 M Mode
6.4.1 M Mode Image Scanning
Perform the following procedure:
1. Select a high-quality image during B mode scanning, and adjust to position the area of interest
in the center of the B mode image.
2. Tap [M] to enter M sampling line status, and drag the sampling line to the desired position.
3. Tap [M]/ [Update] or double-tap the sampling line to enter M mode. You can then observe the
tissue motion along with the anatomical images of B mode. During the scanning process, you
can also adjust the sampling line accordingly when necessary.
4. Tap [Image] to open the image menu. Adjust the parameters to optimize the image.

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals