6 - 34 Operator’s Manual
6 Image Acquisition
2. Select [Smart] > [Auto GA], the system will automatically recognize and trace the target
boundary to calculate.
The calculation result is displayed on the screen.
3. If the calculation result is not satisfactory, adjust the trace to recalculate.
a. Tap the traced contour to activate the cursor.
b. Tap to anchor the desired point on the traced contour.
c. Tap and hold the hand icon to move the cursor to the desired position, you can also tap the
arrows around the hand icon to fine-tune the cursor position.
d. Repeat steps b to c above to adjust more points if needed.
The calculation results display in real time.
4. Tap [Accept Results] to accept the result.
After accepting the result, if the accepted result is not satisfactory, tap [Remeasure] and repeat
step 3 above to update the result.
5. Select [Trend Curve] to check the trend curve of all the accepted results.
6.16 Auto DFR
This function is used to evaluate the diastole function of the left ventricular. E/A and E/ E' are the
indexes of diastolic function and can be automatically measured.
Perform the following procedure:
1. Select a phased probe (except TEE probes) and cardiac exam mode.
2. Scan the apical 4 chamber views.
During scanning, adjust the imaging parameters to optimize the image.
3. Tap [Smart] > [Auto DFR] to enter this function.
If necessary, adjust the position of the sampling volume.
4. Select the desired measurement item (MV E/A, MV E/E' Septal, MV E/E' Lateral), and the
calculation results are automatically displayed in the upper right corner of the screen.
5. Tap [Accept Result] to confirm and save the calculation results.
If necessary, adjust the position of E or A peak point to fine tune the calculation result, and
then select [Accept Result].
Measurement results include Peak E-wave velocity, Peak A-wave velocity, MV E/A ratio,
Pulsed-wave TDI E' velocity and Mitral E/E' ratio.
6. Tap [Exit] to end the calculation and exit Auto DFR.
6.17 Smart Echovue
This function is to support training in recognizing the standard view type, as well as detect and
display the feature structure of the standard view for the cardiac ultrasound image.
Phased probe in cardiac exam mode (except for neonatal cardiac), single B mode, or real time/
freeze/cine mode support this function.

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals