1 - 6 Operator’s Manual
1 Important Information
• A transducer sheath must be installed over the transducer before
performing examination.
• When using a probe, pay attention to the status of the ultrasound image.
Do not use the probe to perform image acquisition when the image is
frozen.
CAUTION
• When using the probe, wear sterile gloves to prevent infection.
• Please use the ultrasound gel com-pliant with the relevant local
regulations. And manage the ultrasound gel properly to ensure that it does
not become a source of infection.
• In normal diagnostic ultrasound mode, there is no danger of a normal-
temperature burn; however, do not keep the probe on the same region of
the patient for more than 10 minutes to avoid risk of burn.
• Do not use the carrying case for storing the transducer. If the carrying case
is used for storage, it may become a source of infection.
• It is required to practice ALARA when operating ultrasound system.
Minimize the acoustic power without compromising the quality of images.
• The probe and accessories supplied with it are not delivered disinfected or
sterilized. Sterilization (or high-level disinfect) before use is required.
• Disposable components should be packaged sterile and for single-use
only. Do not use if integrity of packaging violated or if expiration date has
passed. Please use the disposable components compliant with the relevant
local regulations.
• Please use the disinfection or sterilization solution recommended in this
operator's manual; otherwise Mindray will not be liable for damage caused
by other solutions. If you have any questions, please contact Mindray
Customer Service Department or sales representative.
• Do not use pre-lubricated condoms as a sheath. Lubricant may not be
compatible with the probe material and damage may result.
• The damage of the transducer may be caused by the contact of improper
gel or cleaner:
– DO NOT dip the transducer in the strong polar solution of ethanol,
chloride of lime, ammonium chloride, acetone and formaldehyde.
– DO NOT contact the transducer with solution or ultrasound gel
containing oily medium such as mineral oil or lanoline.
NOTE:
• Read the following precautions to prevent the probe from malfunction:
– Before connecting or disconnecting the probe, freeze or turn off the system.
– Clean and disinfect the probe before and after each examination.

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals