Operating basics
RUN. Runs the di
agnostic routines for the mainframe and the installed modules.
The instrument must be in Factory mode in order to access this menu selection.
CAUTION. Before you execute the diagnostic routines, be sure to save the current
instrument settings to a preset. After the diagnostics are completed, reboot the
instrument and then recall the saved settings. You do not need to re boot the
instrument if you run the diagnostics while the instrument is in normal mode.
Press the ENTER button to execute the diagnostic routines.
The GPS7 module has its own diagnostics submenu; running the mainframe
diagnostics here does not run the GPS7 diagnostics. (See page 3-152, GPS7
module DIAGNOSTICS submenu.) If all of the diagnostic routines are completed
without
any errors, the instrument will display "Pass".
If a diagnostic error is displayed, contact your local Tektronix Field Office or
represe
ntative.
NOTE. The front-panel buttons are disabled until the diagnostic routines are
completed.
2–46 TG8000 Multiformat Test Signal Generator User Manual