Models 2500 and 2502 User’s Manual Range, Digits, Speed, and Filters 6-13
Remote filter programming
Filter commands
Table 6-6 summarizes filter commands. See Section 17 for more details.
* Since filter commands are global, :SENSe[1] (channel 1) commands also affect equivalent :SENSe2 (channel 2) settings.
Table 6-6
Filter commands
Commands* Description
for Average Filter:
[:SENSe[1]]:AVERage:COUNt <n>
[:SENSe[1]]:AVERage:TCONtrol <name>
[:SENSe[1]]:AVERage[:STATe] <state>
[:SENSe[1]]:AVERage:ADVanced:NTOLerance <NRf>
[:SENSe[1]]:AVERage:ADVanced[:STATe] <state>
:SENSe2:AVERage:COUNt <n>
:SENSe2:AVERage:TCONtrol <name>
:SENSe2:AVERage[:STATe] <state>
:SENSe2:AVERage:ADVanced:NTOLerance <NRf>
:SENSe2:AVERage:ADVanced[:STATe] <state>
for Median Filter:
[:SENSe[1]]:MEDian:RANK <NRf>
[:SENSe[1]]:MEDian[:STATe] <state>
:SENSe2:MEDian:RANK <NRf>
:SENSe2:MEDian[:STATe] <state>
Set Ch. 1 average filter count (n = count, 1 to
100).
Select Ch. 1 filter type (name = MOVing or
REPeat).
Enable/disable Ch. 1 filter (state = ON or OFF).
Set Ch. 1 advanced filter noise window in %
(NRf = noise window, 0 to 105).
Enable/disable Ch. 1 advanced filter (state = ON
or OFF).
Set Ch. 2 average filter count (n = count, 1 to
100).
Select Ch. 2 filter type (name = MOVing or
REPeat).
Enable/disable Ch. 2 filter (state = ON or OFF).
Set Ch. 2 advanced filter noise window in %
(NRf = noise window, 0 to 105).
Enable/disable Ch. 2 advanced filter (state = ON
or OFF).
Set Ch. 1 median filter rank (NRf = rank, 0 to 5).
Enable/disable Ch. 1 median filter (state = ON or
OFF).
Set Ch. 2 median filter rank (NRf = rank, 0 to 5).
Enable/disable Ch. 2 median filter (state = ON or
OFF).