Model 2500 and 2502 User’s Manual Getting Started 1-15
SWEEP — Select this option if you want the display to disable while performing a sweep.
The display will disable as soon as sweep is started. The display will automatically re-
enable after the sweep is completed.
STORE — Select this option if you want the display to disable when storing source-
measure readings in the buffer. The display will disable as soon as the buffer is enabled.
The display will automatically re-enable after the storage process is completed. Note that
with this option, the display will disable while performing a sweep. Sweep readings are
automatically stored in the buffer.
Remote display programming
The display can be controlled by various SCPI :DISPlay subsystem commands. Table 1-2
summarizes basic commands. See Section 17, “DISPlay subsystem, for more information
on using these commands.
Front panel tests
Use the TEST/DISPLAY TESTS selection of the main MENU to test various aspects of
the front panel. Test selections include:
KEYS — Front panel keys are tested. Pressing a key displays a message that identifies
that key. Pressing EXIT twice cancels this test.
DISPLAY PATTERNS — Use this selection to turn on all display pixels and annun-
ciators. Subsequent key presses cycle through tests that turn off annunciators and cor-
ner pixels of each digit, turn on the rows of the top-left display digit, and turn on all
annunciators and pixels of each digit in a sequential manner. Press EXIT to cancel this
test.
CHAR SET — This test displays special characters. Press EXIT to cancel the test.
See “Menus,” page 1-20, for more menu information.
Table 1-2
Basic display commands
Command Description
:DISPplay:ENABle <state>
:DISPlay:DIGits <n>
:DISPlay:MODE<name>
Enable/disable display (state = ON or OFF).
Set display resolution (n = 3 to 6).
Select display mode (name = CALC3,
CALC4, CALC5, CALC6, or DUAL).
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2500 and is the answer not in the manual?

Keithley 2500 Specifications

General IconGeneral
BrandKeithley
Model2500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals