Models 2500 and 2502 User’s Manual Status and Error Messages B-9
-420, Query UNTERMINATED
This error occurs when you address the instrument to talk, and there is no response mes-
sage to send. The most likely causes are:
Not sending a query You must send a valid query to the instrument before address-
ing it to talk.
Sending an invalid query If you have sent a query and still get this error, make sure
that the instrument is processing the query without error. For example, sending an ill-
formed query that generates an error -113, Undefined header and then addressing the
instrument to talk will generate an error -420, Query UNTERMINATED as well.
•Valid query following an invalid commandThis situation can occur when you send
multiple commands or queries (program message units) within one command string
(program message). When the Model 2500 detects an error in a program message unit,
it discards all further program message units until the end of the string; for example:
:SENS1:DATE?; :SOUR1:VOLT?
In the above program message, the program message unit :SENS1:DATE? will gener-
ate error -113, Undefined header, and the Model 2500 will discard the second program
message unit :SOUR1:VOLT? even though it is a valid query.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2500 and is the answer not in the manual?

Keithley 2500 Specifications

General IconGeneral
BrandKeithley
Model2500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals