6-14 Range, Digits, Speed, and Filters Models 2500 and 2502 User’s Manual
Filter programming example
Table 6-7 summarizes the command sequence to program channel 1 filter aspects as
follows:
•Average filter off
Median filter state: on
Median filter rank: 5
•Average filter state: on
•Average filter type: moving
•Average filter count 20
Advanced filter state: on
Advance filter tolerance: 10%
Table 6-7
Filter programming example
Command Description
:MED:RANK 5
:MED ON
:AVER:TCON MOV
:AVER:COUN 20
:AVER ON
:AVER:ADV OFF
:AVER:ADV:NTOL 10
Set median rank to 5.
Enable median filter.
Select moving filter type.
Set moving count to 20.
Enable average filter.
Disable advanced filter.
Set filter tolerance to 10%.