Models 2500 and 2502 User’s Manual Sweep Operation 9-5
Custom sweep examples
The custom sweep can be configured to provide a 50% duty cycle pulse sweep. Figure 9-3
shows a pulse sweep that provides three 1V pulses on a 0V bias level. This pulse sweep is
configured by specifying six points for the custom sweep. The specified voltage levels at
points P0, P2, and P4 are 1V, and the specified voltage levels at points P1, P3, and P5 are
0V. Six measurements are performed for this sweep, three at 1V and three at 0V.
Figure 9-3
Custom pulse sweep
Figure 9-4 shows a custom sweep example with different pulse widths. In this example,
the first two points are configured with the same source value so that the duration of the
first pulse is effectively doubled.
Figure 9-4
Custom sweep with different pulse widths
Delay
P1
Measure
#1
P0
Measure
#2
Delay
Measure
#3
Measure
#4
Measure
#5
Measure
#6
Bias 0V
1V
Bias
Delay
Delay
P3
Delay
P5
Delay
P2 P4
1V
0V
#1 #2
#3 #4
#5
#6 #7
Delay Delay Delay Delay Delay Delay
All have same delay, but pulse widths differ.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2500 and is the answer not in the manual?

Keithley 2500 Specifications

General IconGeneral
BrandKeithley
Model2500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals