2-10 Connections Model 2500 and 2502 User’s Manual
Voltage source connections
Figure 2-8 shows the connecting scheme for using the Model 2500 as a stand-alone volt-
age source. Note that the DUT HI terminal is connect to the OUTPUT terminal for the
voltage source, while the DUT LO terminal is connected to analog common (inner ring of
INPUT jack). The maximum voltage source output is 100V @ 20mA.
Figure 2-8
Stand-alone voltage source connections
WARNING: NO INTERNAL OPERATOR SERVICABLE PARTS, SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION: FOR CONTINUED PROTECTION AGAINST FIRE HAZARD, REPLACE FUSE WITH SAME TYPE AND RATING.
120
LINE RATING
50, 60Hz
60 VA MAX
FUSE LINE
630 mAT
(SB)
315 mAT
(SB)
100 VAC
120 VAC
220 VAC
240 VAC
CAT I
MADE IN
U.S.A.
DIGITAL I/O
RS-232
TRIGGER LINK
INPUT
CHANNEL 1
INPUT
CHANNEL 2
VOLTAGE SOURCE
OUTPUT CHANNEL 1
VOLTAGE SOURCE
OUTPUT CHANNEL 2
RATINGS MAX
100V @ 20mA
COMMON
MODE
200V
RATINGS MAX
100V @ 20mA
!
!
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
Model 2500
INPUT LO
(Inner Shield)
Optional Noise Shield
(Connect to INPUT LO)
Channel 1 INPUT
Triax Cable
DUT
Channel 1 OUTPUT
LO
HI
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2500 and is the answer not in the manual?

Keithley 2500 Specifications

General IconGeneral
BrandKeithley
Model2500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals