17-38 SCPI Command Reference Models 2500 and 2502 User’s Manual
MODE <name>
:CALCulate7:CLIMits:MODE <name> Control digital I/O port pass/fail output
Parameters <name> = GRADing Output graded pass/fail pattern
SORTing Output sorted pass/fail pattern
Query :MODE? Query digital I/O pass/fail mode
Description This command controls how limit calculations drive the digital I/O
lines. In GRADing mode, a reading passes if it is within all of the hi/low
limit tolerances enabled, assuming that it has passed Limit 1 or Limit 2
compliance test first. The digital I/O lines will be driven with the first
pattern of the first compliance, hi, or low failure. Otherwise, the
CALC7:CLIM:PASS:SOUR3 pattern will be output.
In SORTing mode, a reading will fail if it fails the compliance test, or is
not within any of the digital I/O bands. If the tests pass and only Limit 1
or Limit 2 is enabled, the CALC7:CLIM:PASS:SOUR7 pattern will be
output. Otherwise, the first limit test band that passes will output its
LOW:SOUR3 pattern (UPP:SOUR3 patterns will be ignored). If Limit 1
or Limit 2 fails, its SOUR3 pattern will be output. If no Limit 3-6 limit
passes, the CALC7:CLIM:FAIL:SOUR3 pattern will be output.
Clear test results
[:IMMediate]
:CALCulate7:CLIMits:CLEar[:IMMediate] Clears test results and resets digital I/O
port
Description This command clears the test results (pass or fail) of the limit tests and
resets the output lines of the digital I/O port back to the :SOURce3:TTL
settings. See “SOURce3,” page 17-74.
AUTO <b>
:CALCulate7:CLIMits:CLEar:AUTO <b> Control auto-clear for test results
Parameters <b> = 1 or ON Enable auto-clear
0 or OFF Disable auto-clear
Query :AUTO? Query state of auto-clear
Description With auto-clear enabled, test results will clear and the output lines of the
digital I/O port will reset when the :INITiate command is sent to start a
new test sequence.
When disabled, you must use :IMMediate to perform the clear actions.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2500 and is the answer not in the manual?

Keithley 2500 Specifications

General IconGeneral
BrandKeithley
Model2500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals