10-12 Triggering Models 2500 and 2502 User’s Manual
Switching mainframe setup
Step 1. Restore bench defaults.
Press the MENU key, select SAVESETUP, and then press ENTER. From the
SAVESETUP menu, select RESET, then press ENTER. Press ENTER to confirm the
action. Press ENTER to return to the SETUP MENU. Press EXIT to leave the SETUP
MENU. Press EXIT to leave the MAIN MENU.
Step 2. Set up scan list: 1!1 - 1!10.
Press the SCAN LIST key. Press 1, 1 - 1, 1, 0, then press the ENTER key.
Step 3. Set the number of scans to 1.
Press the SCAN key, select SCAN-CONTROL, and then press ENTER. From the SCAN
CONTROL menu, select NUMBER-OF-SCANS, then press ENTER. Select ENTER-
SCAN-COUNT, then press ENTER. Set the count to 1, then press ENTER. Press the
EXIT key to return to the CONFIGURE SCAN menu.
Step 4. Set channel spacing to trigger-link.
Select CHAN-CONTROL from the CONFIGURE SCAN menu, then press ENTER.
Select CHANNEL-SPACING from the CHANNEL CONTROL menu, then press
ENTER. Select TRIGLINK, then press ENTER. Select ASYNCHRONOUS, then press
ENTER. Press ENTER, ENTER, then EXIT, EXIT, EXIT to leave the CONFIGURE
SCAN menu.
Operation
1. To store the readings in the Model 2500 buffer, press STORE, and set the buffer size
for 10. When ENTER is pressed, the asterisk (*) annunciator will turn on to indicate
the buffer is enabled. See Section 8 for details.
2. Turn the Model 2500 OUTPUT ON. The Model 2500 waits for an external trigger
from the switching mainframe.
3. Press STEP on the Model 7001/7002 to take it out of idle and start the scan. The scan-
ner's output pulse triggers the Model 2500 to take a reading and store it. The Model
2500 then sends a trigger pulse to the switching mainframe to close the next channel.
This process continues until all 10 channels are scanned, measured, and stored.
Details of this testing process are explained in the following paragraphs and are referenced
to the operation model shown in Figure 10-7.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2500 and is the answer not in the manual?

Keithley 2500 Specifications

General IconGeneral
BrandKeithley
Model2500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals