Models 2500 and 2502 User’s Manual Common Commands 15-5
*SAV, *RCL programming example
Table 15-3 summarizes the basic command sequence for saving and recalling a setup. The
present setup is stored in memory location 2, GPIB defaults are restored, and the memory
location 2 setup is recalled.
*RST — reset Return Model 2500 to GPIB defaults
When the *RST command is sent, the Model 2500 performs the following operations:
• Returns the Model 2500 to the GPIB default conditions. Refer to “Default parameters”
column of SCPI tables in Section 17.
• Cancels all pending commands.
• Cancels response to any previously received *OPC and *OPC? commands.
*TRG — trigger Send bus trigger to Model 2500
Use the *TRG command to issue a GPIB trigger to the Model 2500. It has the same effect
as a group execute trigger (GET).
Use the *TRG command as an event to control operation. The Model 2500 reacts to this
trigger if BUS is the programmed arm control source. The control source is programmed
from the TRIGger subsystem.
NOTE Details on triggering are covered in Section 10.
Table 15-3
*SAV, *RCL programming example
Command Description
*SAV 2
*RST
*RCL 2
Save present setup in memory location 2.
Restore GPIB defaults.
Recall location 2 setup.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com