17-74 SCPI Command Reference Models 2500 and 2502 User’s Manual
SOURce3
The following commands are used to set the logic level of the digital output lines, and con-
trol the pulse width of limit test output patterns that are sent to component handlers. Limit
tests are configured and controlled from the CALCulate7 Subsystem. For details on limit
testing, see Section 11.
Setting digital output
[:LEVel] <NRf> | <NDN>
:SOURce3:TTL[:LEVel] <NRf> | <NDN> Set digital output pattern
Parameters <NRf> = 0 to 7 Decimal format for 3-bit
0 to 15 Decimal format for 4-bit
<NDN> #Bx Binary format:
3-bit: x = 000 to 111
4-bit: x = 0000 to 1111
#Hx Hexadecimal format:
3-bit: x = 0 to 7
4-bit: x = 0 to F
#Qx Octal format:
3-bit: x = 0 to 7
4-bit: x = 0 to 17
Query :TTL? Query digital output value
Description This command is used to set the logic levels of the output lines of the
digital I/O port. When set high, the specified output line will be at
approximately +5V. When set low, the output line will be at 0V.
NOTE The maximum digital output value is 65,535 with the 2499-DIGIO option.