10-24 Triggering Models 2500 and 2502 User’s Manual
Remote trigger commands
Table 10-1 summarizes remote trigger commands. These commands are covered in more
detail in Section 17, “Trigger subsystem,” except for *TRG, a common command covered
in Section 15.
Table 10-1
Remote trigger command
Command Description
:INITiate
:ABORt
:ARM:COUNt <n>
:ARM:SOURce <name>
:ARM:TIMer <n>
:ARM:DIRection <name>
:ARM:ILINe <NRf>
:ARM:OLINe <NRf>
:ARM:OUTPut <event list>
:TRIGger:CLEar
:TRIGger:COUNt <n>
:TRIGger:DELay <n>
:TRIGger:SOURce <name>
:TRIGger:DIRection <name>
:TRIGger:ILINe <NRf>
:TRIGger:OLINe <NRf>
:TRIGger:INPut <event list>
:TRIGger:OUTPut <event list>
*TRG
Take Model 2500 out of idle state.
Abort operation, return to idle.
Set arm count (n = count).
Specify arm control source. (Name = IMMediate, TLINk, TIMer,
MANual, BUS, NSTest, PSTest, or BSTest.)
Set arm layer timer interval (n = interval).
Control arm bypass. (Name = SOURce or ACCeptor.)
Select arm layer input line (NRf = input line #).
Select arm layer output line (NRf = output line #).
Select arm layer output events. (Event list = TRIGger or NONE.)
Clear any pending input triggers immediately.
Set trigger count (n = count).
Set trigger delay (n = delay).
Specify trigger control source. (Name = IMMediate or TLINk.)
Control trigger bypass. (Name = SOURce or ACCeptor.)
Select trigger layer input line (NRf = input line).
Select trigger layer output line (NRf = output line).
Select trigger input layer events. (Event list = SOURce, DELay,
SENSe, or NONE.)
Select trigger layer output events. (Event list = SOURce, DELay,
SENSe, or NONE.)
Trigger Model 2500 (if BUS source selected).
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176