Models 2500 and 2502 User’s Manual Photodiode Measurements 4-7
Photodiode measurement programming example
Table 4-3 summarizes the command sequence for a photodiode measurement procedure
using channel 1 to measure the current of the back photodiode detector, and channel 2
displaying optical power. Note that the steps correspond to those listed previously in
“Front panel photodiode measurement procedure.” See Figure 4-2 earlier in this section
for basic connections and Section 2 for detailed connection information.
These commands set up the Model 2500 as follows:
• Channel 1 measurement: current measurement, auto range
• Channel 2 measurement: optical power
• Channel 2 responsivity constant (R) of detector for wavelength of interest: 1
• Channel 2 dark current (D): 1
• Channel 1 source: 10V range, +10V amplitude
• Channel 2 source: 100V range, +20V amplitude
NOTE See Appendix H for a complete program listing.