9-6 Sweep Operation Models 2500 and 2502 User’s Manual
Configuring and running a sweep
Front panel sweep operation
Configuring a sweep
The sweep configuration menu is structured as follows and shown in Figure 9-5. Note that
bullets indicate the primary items of the sweep menu and dashes indicate the options of
each menu item. Using Section 1, “Rules to navigate menus,” go through the following
menu to select and configure the desired sweep.
NOTE Before configuring a sweep, press SRC1 or SRC2 to select the channel to be
used for the sweep. You can also switch between channel 1 and channel 2 by
pressing SRC1 or SRC2 while in the CONFIGURE SWEEPS menu.
CONFIGURE SWEEPS menu:
Press CONFIG then SWEEP to display the sweep configuration menu.
• TYPE — Use this menu item to select the type of sweep:
– NONE – Disables all sweeps.
–STAIR – When the linear staircase sweep is selected, you will be prompted to enter
the START, STOP, and STEP levels.
– LOG – When the logarithmic staircase sweep is selected, you will be prompted to
enter the START and STOP levels and specify the number of measurement points.
– CUSTOM – With the custom sweep selected, you specify the number of measure-
ment points (#-POINTS) in the sweep and the source level at each point (ADJUST-
POINTS). With the INIT option, you can set a consecutive range of measurement
points in the sweep to a specific level. For example, assume that for a 20-point cus-
tom voltage sweep (# POINTS = 20), you want points 10 through 15 to be set for
1V. After selecting the INIT option, set the VALUE to +1.000000V, set the START
PT to 10, and set the STOP PT to 15.
• SWEEP-COUNT — Use this menu item to specify how many sweeps to perform:
– FINITE – Use this option to enter a discrete number of sweeps to perform with the
results stored in the data store buffer. The maximum number of finite sweeps that
can be performed is determined as follows:
maximum finite sweep count = 3000 / # Points in sweep
– INFINITE – Select this option to continuously repeat the configured sweep. Use
the EXIT key to stop the sweep. Data is not stored in the buffer.