Models 2500 and 2502 User’s Manual SCPI Command Reference 17-95
ILINe <NRf>
:ARM[:SEQuence[1]][LAYer[1]][:TCONfigure]:ILINe <NRf> Select input line; arm layer
:TRIGger[:SEQuence[1]][:TCONfigure]:ILINe <NRf> Select input line; trigger layer
Parameters <NRf> = 1 Line #1
2 Line #2
3 Line #3
4 Line #4
5 Line #5
6 Line #6
Query :ILINe? Query input trigger line
Description This command is used to select input lines for the Trigger Link. For nor-
mal operation, Trigger Link input and output (“OLINe”) should not
share the same line.
OLINe <NRf>
:ARM[:SEQuence[1]][LAYer[1]][:TCONfigure]:OLINe <NRf> Select output line; arm layer
:TRIGger[:SEQuence[1]][:TCONfigure]:OLINe <NRf> Select output line; trigger layer
Parameters <NRf> = 1 Line #1
2 Line #2
3 Line #3
4 Line #4
5 Line #5
6 Line #6
Query :OLINe? Query output trigger line
Description This command is used to select output lines for the Trigger Link. For
normal operation, Trigger Link input and output (“ILINe”) should not
share the same line.