Models 2500 and 2502 User’s Manual SCPI Command Reference 17-89
CONTrol <name>
:TRACe:FEED:CONTrol <name> Start or stop buffer
Parameters <name> = NEXT Fills buffer and stops
NEVer Disables buffer storage
Query :CONTrol? Query buffer control
Description This command is used to select the buffer control. When NEXT is
selected, the asterisk (*) annunciator turns on to indicate that the buffer
is enabled. The storage process starts when Model 2500 is taken out of
idle to perform source-measure operations.
After the buffer stores the specified number of reading arrays (as set by
the :POINts command), the asterisk annuciator turns off to indicate that
storage is done.
With NEVer selected, storage into the buffer is disabled.
Select timestamp format
FORMat <name>
:TRACe:TSTamp:FORMat <name> Select timestamp format
Parameters <name> = ABSolute Reference to first buffer reading
DELTa Time between buffer readings
Query :FORMat? Query timestamp format
Description This command is used to select the timestamp format for buffer read-
ings. With ABSolute selected, each timestamp is referenced to the first
reading stored in the buffer. With DELTa selected, timestamps provide
the time between each buffer reading.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2500 and is the answer not in the manual?

Keithley 2500 Specifications

General IconGeneral
BrandKeithley
Model2500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals