Models 2500 and 2502 User’s Manual SCPI Command Reference 17-53
Output enable control
ENABle[:STATe] <b>
:OUTPut[1]:ENABle[:STATe] <b> Control output enable for both sources
Parameters <b> = 0 or OFF Disable output enable
1 or ON Enable output enable
Query :ENABle[:STATe]? Query state of output enable
Description This command is used to enable or disable the output enable line. When
enabled, the sources cannot be turned on unless the output enable line
(pin 8 of the rear panel digital I/O connector) is pulled to a logic low
state. When the output enable line goes to a logic high state, both
sources turn off. See Section 12, “Digital I/O port” and “Output enable,”
for details using output enable with a test fixture.
When disabled, the logic level on the output enable line has no effect on
the output state of the source.
TRIPped?
:OUTPut[1]:ENABle:TRIPped?
Description This query command is used to determine if output enable has been
tripped. The tripped condition (“1”) means that the sources can be
turned on (output enable line at logic low level).
A “0” will be returned if the sources cannot be turned on (output enable
line at logic high level).