7-10 Relative, Math, Ratio, and Delta Models 2500 and 2502 User’s Manual
Front panel RATIO and DELTA configuration
Using Table 7-6 as a guide, configure RATIO and DELTA functions as follows:
1. Press CONFIG then RATIO or DELTA to select the function to be configured.
2. Using the EDIT keys, place the cursor on the desired function, then press ENTER.
3. From normal display, press the RATIO or DELTA key to display the desired function
reading.
4. To cancel the function and return to normal display, press the MSR1 or MSR2 key.
Table 7-6
RATIO and DELTA configuration menus
Configuration menu item Description
CONFIG RATIO
CONFIG RATIO BUTTON
MSR1/MSR2
MSR2/MSR1
CONFIG DELTA
CONFIG DELTA BUTTON
MSR1-MSR2
MSR2-MSR1
Configure RATIO function.
MSR1/MSR2 function.
MSR2/MSR1 function.
Configure DELTA function.
MSR1-MSR2 function.
MSR2-MSR1 function.