17-64 SCPI Command Reference Models 2500 and 2502 User’s Manual
TRIGgered[:AMPLitude] <n>
:SOURce[1]:VOLTage[:LEVel]:TRIGgered[:AMPLitude] <n> Set fixed source 1 amplitude when
triggered
:SOURce2:VOLTage[:LEVel]:TRIGgered[:AMPLitude] <n> Set fixed source 2 amplitude when
triggered
Parameters <n> = -100 to 100 Set source amplitude (volts)
DEFault 0V
MINimum -100V
MAXimum +100V
Query :TRIGgered? Query triggered amplitude for fixed source
:TRIGgered? DEFault Query *RST default amplitude
:TRIGgered? MINimum Query lowest allowable amplitude
:TRIGgered? MAXimum Query highest allowable amplitude
Description These commands perform the same as the [:IMMediate][:AMPLitude]
commands except that the source amplitude is not updated immediately.
With this command, the amplitude is updated when the Model 2500 is
triggered to perform a source and measure operation. For example, if the
instrument is waiting in the trigger layer for an external trigger, the
amplitude of the source will not update until that external trigger is
received by the Model 2500. See Section 10, “Triggering,” for details on
trigger model operation.
The MINimum and MAXimum parameters are only valid if the 100V
source range is presently selected. Sending the MINimum or MAXi-
mum parameters on the 10V source range will generate error -221 (Set-
ting Conflict).