Model 2500 and 2502 User’s Manual Connections 2-11
Analog output connections (Model 2502 only)
Analog output connector terminals
Figure 2-9 shows the terminal configuration of each triax ANALOG OUT connector.
Output terminals are as follows:
• Center conductor: analog output HI
• Inner shield: analog output LO (floating analog common)
• Outer shield: chassis ground
WARNING Analog output low can float up to ±100V. Use care to avoid a shock
hazard when using the analog outputs.
Figure 2-9
Analog output connector terminals
Analog Output HI
Analog Output LO
(Analog Common)
Chassis Ground
ANALOG OUT
Triax Connector