14-18 Status Structure Models 2500 and 2502 User’s Manual
Event enable registers
As Figure 14-1 shows, each status register set has an enable register. Each event register
bit is logically ANDed (&) to a corresponding enable bit of an enable register. Therefore,
when an event bit is set and the corresponding enable bit is set (as programmed by the
user), the output (summary) of the register will set to 1, which in turn sets the summary bit
of the status byte register.
The commands to program and read the event enable registers are listed in Table 14-7. For
details on programming and reading registers, see “Programming enable registers,
page 14-5 and “Reading registers,” page 14-6.
NOTE The bits of any enable register can be reset to 0 by sending the 0 parameter
value with the appropriate enable command (i.e. STATus:OPERation:ENABle
0).
Table 14-7
Event enable register commands
Command Description Default
*ESE <NDN> or <NRf>
*ESE?
STATus
:OPERation
:ENABle <NDN> or <NRf>
:ENABle?
:MEASurement
:ENABle <NDN> or <NRf>
:ENABle?
:QUEStionable
:ENABle <NDN> or <NRf>
:ENABle?
Program Standard Event Enable Register. See “Parameters.
Read Standard Event Enable Register.
STATus Subsystem:
Operation Event Enable Register:
Program enable register. See “Parameters.
Read enable register.
Measurement Event Enable Register:
Program enable register. See “Parameters.
Read enable register.
Questionable Event Enable Register:
Program enable register. See “Parameters.
Read Measurement Event Enable Register:
Note
Parameters:
<NDN> = #Bxx…x Binary format (each x = 1 or 0)
= #Hx Hexadecimal format (x = 0 to FFFF)
= #Qx Octal format (x = 0 to 177777)
<NRf> = 0 to 65535 Decimal format
Note: Power-up and STATus:PRESet resets all bits of all enable registers to 0. *CLS has no effect.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2500 and is the answer not in the manual?

Keithley 2500 Specifications

General IconGeneral
BrandKeithley
Model2500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals