17-48 SCPI Command Reference Models 2500 and 2502 User’s Manual
Grading mode status bit values:
Sorting mode status bit values:
Result Bit #: 12 11 10 9 8
Meas.
Event
Status
1
All limit tests passed 0 0 0 0 0 Bit 5 (LP)
Limit test 1 failed 0 0 0 0 1 Bit 0 (L1)
Limit test 2 failed 0 0 0 1 0 Bit 1 (L2)
Limit test 3 high failed 1 0 0 1 1 Bit 4 (LFH)
Limit test 3 low failed 0 0 0 1 1 Bit 2 (L3)
Limit test 4 high failed 1 0 1 0 0 Bit 4 (LFH)
Limit test 4 low failed 0 0 1 0 0 Bit 3 (L4)
Limit test 5 high failed 1 0 1 0 1 -
Limit test 5 low failed 0 0 1 0 1 -
Limit test 6 high failed 1 0 1 1 0 Bit 4 (LFH)
Limit test 6 low failed 0 0 1 1 0 -
1
See Section 14, “Measurement event register, and Figure 14-6 for details.
Result Bit #: 12 11 10 9 8
Meas.
Event
Status
1
Limit 1 & 2 passed, 3-6 disabled 0 0 0 0 0 Bit 5 (LP)
Limit test 1 failed 0 0 0 0 1 Bit 0 (L1)
Limit test 2 failed 0 0 0 1 0 Bit 1 (L2)
Limit test 3 passed 0 0 0 1 1 Bit 3 (L3)
Limit test 4 passed 0 0 1 0 0 Bit 4 (L4)
Limit test 5 passed 0 0 1 0 1 -
Limit test 6 passed 0 0 1 1 0 -
All Limits 3 - 6 failed,
Limits 1 and 2 passed
1 1 1 1 1 -
1
See Section 14, “Measurement event register, and Figure 14-6 for details.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2500 and is the answer not in the manual?

Keithley 2500 Specifications

General IconGeneral
BrandKeithley
Model2500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals