17-6 SCPI Command Reference Models 2500 and 2502 User’s Manual
Command Description
Default
parameter SCPI
:CALCulate7
:LIMitX
:UPPer
[:DATA] <n>
[:DATA]?
:SOURce3 <NRf> | <NDN>
:SOURce3?
:LOWer
[:DATA] <n>
[:DATA]?
:SOURce3 <NRf> | <NDN>
:SOURce3?
:PASS
:SOURce3 <NRf> | <NDN>
:SOURce3?
:STATe <b>
:STATe?
:FAIL?
:CLIMits
:BCONtrol <name>
:BCONtrol?
:MODE <name>
:MODE?
CALC7 limit tests (continues):
Path to software limits. LIMitX = LIMit3...LIMit6.
3
Configure upper limit:
Specify upper limit (-9.999999e20
to 9.999999e20).
Query upper limit.
Specify grading mode fail pattern (0 to 7 | 15).
1
Query fail bit pattern.
2
Configure lower limit:
Specify lower limit (-9.999999e20 to
9.999999e20).
Query lower limit.
Specify grading mode fail pattern (0 to 7 | 15).
1
Query fail bit pattern.
2
Path to sorting mode pass:
Specify output pass pattern (0 to 7 | 15).
1
Query pass bit pattern.
2
Enable or disable Limit X test.
Query state of Limit X test.
Return result of Limit X test: 0 (pass) or 1 (fail).
Composite limits for Limit 1 through Limit 6:
Specify when to send binning info to handler. A
limit test is performed (IMMediate) or after a
sweep, list, or memory sequence (END).
Query binning control.
Set how limit results control Digital I/O lines
(GRADing or SORTing).
Query limit results control of Digital I/O lines.
1
15 or 7
-1
15 or 7
15 or 7
OFF
IMM
GRAD
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
1
Default based on present digital output size (:SOURce3:BSIZe <n>). 3-bit default is 7. 4-bit default is 15. 16-bit default is 65535.
(16-bit available only with 2499-DIGIO option.)
2
The format (ASCII, hexadecimal, octal, or binary) for the returned value is set by FORMat:SOURce3 <name>.
3
LIMitX = LIMit3, LIMit4, LIMit5, or LIMit6.
Table 17-1 (continued)
CALCulate command summary