5-22 Image Optimization
Operation
Rotate the knob under [Img Qual.] item or the mapping-menu item on the touch screen
to select the different frequency values.
The adjusting range of frequency values can be divided into 3 levels: penetration
preferred (Pen), general mode (Gen), and resolution preferred (Res).
Select the frequency according to the detection depth and current tissue features.
Effects
The higher the frequency, the better the resolution and sensitivity, and the worse the
force of penetration.
Scale
Description
This function is used to adjust the speed range of the flow, which is adjusted using the
PRF in the system.
The real-time PRF value is displayed in the image parameter area in the top-right
corner of the screen.
Operations
Rotate the knob under the [Scale] item on the touch screen or through the mapping-
menu item.
Effects
Provides a much clearer color flow image.
Use a low PRF to observe low-velocity flows, and a high PRF to observe high-velocity
flows.
Impacts
Aliasing may occur if a low-velocity scale is used and high velocities are encountered.
Low velocities may not be identified when a high-velocity scale is used.
iTouch
Description
To optimize image parameters as per the current tissue characteristics for a better
image effect.
Operations
Press <iTouch> on the control panel to turn the function on.
Auto-Calculation
Description
This function is used to trace the spectrum and calculate the PW/CW mode image
parameters. The results are displayed in the results window.
1. Touch [Auto Calc] on the touch screen or the mapping-menu to turn the auto
calculation function on or off.
2. After the auto calculation function is turned on, touch “Auto Calculation” tab.
Auto
Calculation
Parameter
Select parameters in the dialog box prompted by touching [Auto Calc Param] on the
Auto Calculation page.
Auto
Calculation
Cycle
To set the heart cycle number for auto-calculation.
Touch [Auto Calc Cycle] on the touch screen to select the cycle number, the adjusting
range is 1-5.
Trace Area
To set the trace area of the Doppler wave in the spectrum map, applicable for auto
calculation, V Max and V Mean display.
To change the trace area, adjust through [Trace Area] item or the mapping-menu item
on the touch screen.
The available selections of trace area are: Above, Below, All.
Trace
To set the smooth level when tracing.

Table of Contents

Other manuals for Mindray DC-80A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-80A and is the answer not in the manual?

Mindray DC-80A Specifications

General IconGeneral
BrandMindray
ModelDC-80A
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals