Behavior.
â–
Beep. When Beep is turned on, the instrument beeps when a mask test fails.
â–
E-mail. When E-mail is turned on, the instrument sends an E-mail when a mask test fails. Click the Setup button and
complete the E-mail on Event Setup information.
â–
Save WFM. When Save WFM is turned on, failed waveform data is saved to a file. Click the Path button and enter the path
location for the file.
â–
Log Date. When Log Date is turned on, the date and time for each failed mask test is recorded to a file. Click the Path
button and enter the path location for the file.
â–
Print. When Print is turned on, the instrument sends the waveform to a printer when a mask test fails.
â–
SRQ. When SRQ is turned on, the instrument sends a GPIB SRQ command when a mask test fails.
â–
AUX Out. When AUX Out is turned on, the instrument sends a trigger to the Auxiliary Out connector when a mask test fails.
See External Signals Control Window for additional details on this feature.
What do you want to do next?
Learn about mask polarity.
Learn about waveform parameters.
Go to the Test Setup control window.
Set up mask test completion notification
From the Mask menu, select Pass/Fail Setup.
Overview. Use the Test Completion Notification controls to beep, execute a command, or send an external trigger when a
waveform passes a mask test.
To use. Click a button to switch a control on or off.
Mask testing setups