EasyManuals Logo
Home>Tektronix>Test Equipment>DPO7054C

Tektronix DPO7054C Printable Help

Tektronix DPO7054C
920 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #467 background imageLoading...
Page #467 background image
Behavior. Use the Transition trigger to trigger the instrument on pulse edges that traverse between two thresholds at faster or
slower rates than the specified time. You can set up the instrument to trigger on positive or negative edges. The trigger can also
be qualified by the logical state of other channels or by a bus pattern (MSO70000C/DX Series only). When you click on the
trigger window controls, the trigger graphic shows the trigger sequence.
Use the Upper Level and Lower Level controls to define the voltage ranges of interest. The Time control specifies the target
transition time.
What do you want to do next?
Learn more about other trigger types.
Return to the Trigger Setup control window.
Trigger graphic
The trigger graphic displays a visual indicator of the trigger criteria. Depending on the trigger type, the instrument will update the
graphic as you define the trigger parameters.
Set up a video trigger
NOTE. The video trigger type is available only on DPO7000C, MSO5000B, and DPO5000B Series instruments.
From the Trig menu, select Video Setup.
Overview. Use the controls in this window to set up the Video trigger parameters.
NOTE. The Video trigger is only available on analog channels.
Trigger setups
DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series 435

Table of Contents

Other manuals for Tektronix DPO7054C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO7054C and is the answer not in the manual?

Tektronix DPO7054C Specifications

General IconGeneral
BrandTektronix
ModelDPO7054C
CategoryTest Equipment
LanguageEnglish

Related product manuals