To use.
â–
Click the Trigger Type box and select Bus from the drop-down list.
â–
Click the Bus box and select the bus number or name from the drop-down list.
â–
Click the Logic Thresholds Setup button to set the voltage threshold levels for the channels in the bus.
â–
Select the type of bus cycle or activity to use as the trigger from the Trigger On drop-down list.
â–
For some Trigger On selections, you must set additional fields to define other parameters such as for a Ident & Data.
In this example, you need to set the Trigger When Address is, RT Address type and values, the T/R bit, Subaddress/Mode
type and value, and select the Word Count/ModeCode type and value. You can also set the Parity.
Click the Trigger On selections Status Word, Data, Sync, RT/IMG Time, or Error to view their parameters.
What do you want to do next?
Learn about bus setups.
Trigger on Status Word
Trigger on Data
The error type options available in Trigger on Data are as follows:
Trigger setups