om 5184516-100 Rev. 5 5-57
3-4-9 HU Tone Alert / %HU
This is the Heat Units Remaining level at which the tone is sounded. This is only
followed when VA Site is set to “yes.”
3-4-10 Number of Tubes
(1,2) When set to “1,” only the under table tube is available, all references to Tube
2 are removed.
3-4-11 OTS
(Yes, No) When set to “yes,” table bucky and table top are available however, the
number of tubes must be set to 2.
3-4-12 Wallstand Location
When set to “0”, wall bucky will not be an option as a receptor.
3-4-13 Fluoro Rates
(Continuous = 30, Pulsed = 15, 7.5, 4) These are set values and are options
depending if Pulsed Fluoro is set to yes or no.
3-4-14 Spot Rates
(Single, 1, 2, 3, 4, 5, 6, 7.5) Available record rates.
3-4-15 Max Record rate
(7.5) Fastest record rate
3-4-16 System Log On
(Yes, No) When set to “yes” the system will require user to log on. If set to “no” the
Log out button is not displayed on the utilities screen.