om 5184516-100 Rev. 5 5-24
2-10Emergency Recovery
If during a procedure power is lost to the system, the system will reboot if power is
returned to the unit after 30 seconds.
NOTE: Upon reboot, the system will remember all patient parameters if power was lost
during a procedure. If power is off longer than 30 seconds, the system must be
rebooted by pressing the Power On button located on the RCIM. The study will
not resume. Re-select the patient from the Patient list and the exam.
ILLUSTRATION 5-23
RESET RECOVERY SCREEN
The system will then give the options for continuing the study by selecting
Resume Exam or ending the study by the Close Exam option.
IF THE SYSTEM CONTROL INTEGRATED CONSOLE LOSES
COMMUNICATION WITH THE REMAINING SYSTEM DURING A
DIGITAL RECORD ACQUISITION, THE USER WILL BE ALLOWED TO
CONTINUE/FINISH (EXPOSURE SWITCH IS RELEASED) THE
ACQUISITION. ANY SUBSEQUENT ACQUISITION (FLUORO OR
DIGITAL) WILL BE INHIBITED.