om 5184516-100 Rev. 5 5-3
1-2Tube Warm–Up Screen
NOTE: IMPORTANT!
A tube warm–up should be performed if the unit has not made an exposure in two
hours. This function will help extend the tube life. The Warm Tube button will
appear when a warm–up is required.
Select the Warm Tube button located on the lower right of the screen as shown in
ILLUSTRATION 5-3.
ILLUSTRATION 5-3
WARM TUBE SCREEN
The technical factors as well as the number of exposures are preset. After select-
ing the appropriate tube for warm up, press the exposure button. Maintaining
pressure on the exposure switch will stop at the completion of the warm up proce-
dure. After the soak time has counted to 0 seconds, the close button must be
selected to continue.
Please remember to follow all radiation safety standards
while performing this and all procedures.
WARM TUBE BUTTON
FLUORO TUBE
RAD TUBE
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Precision 500D and is the answer not in the manual?

GE Precision 500D Specifications

General IconGeneral
BrandGE
ModelPrecision 500D
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals