om 5184516-100 Rev. 5 7-33
3-1-1 Precision 500D Table Electrical
Requirements
Electrical Data: Precision 500D Table with Intelligent Digital Device
Ratings: 240VAC, 20 amps, Single phase, 50/60 Hz
Power Supply Requirements:
Generator power unit provides 240VAC, single phase, 50/60 Hz*,
20 amp branch circuit plus ground
Operating Temperature 50 to 100F (+10° to +38°C)
Relative Humidity 20% to 70% (non–condensing)
Altitude 8000 ft. (2438 m) above sea level max.
Storage or Shipping Temperature –40° to 140°F (–40°C to 60°C)
Relative Humidity 20% to 80% (non–condensing)
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).