om 5184516-100 Rev. 5 2-31
NOTE: Remove the table top ground cord for procedures in which the patient is not elec-
trically susceptible, especially for procedures that require table angulation. Cord
removal reduces the chance of damage to equipment from entanglement with the
cord.
Section 6
Safety Summary
The operator must read this manual in its entirety and receive training in system
operation and application prior to operating the system.
DO NOT operate this unit if there is any uncertainty as to the proper functioning of
the system. Refer all servicing to QUALIFIED SERVICE PERSONNEL.
The system has been designed and tested to meet FDA regulations. The following
precautions and procedures must be complied with during all phases of operation
of this instrument, failure to comply could result in injury and/or death to the oper-
ator or the patient. GE Medical Systems assumes no liability for the customer’s
failure to comply.
NOTE: GE Medical Systems suggests that an image quality and system function
check be performed daily, prior to patient data acquisition. This can be done
by acquiring images using a phantom, retrieving them, and then deleting the
file per the operating instructions in this manual.
The system should not be operated such that a patient could come in
contact with any part of the system.
FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE
ORDER OF A PHYSICIAN.
THIS INSTRUMENT MUST BE GROUNDED! A THREE–CONDUCTOR
AC POWER CABLE IS SUPPLIED WITH THIS INSTRUMENT TO
PROVIDE THE PROPER ELECTRICAL GROUND. TO MINIMIZE THE
SHOCK HAZARD, THE MAIN POWER CABLE OF THE ISOLATION
TRANSFORMER MUST BE PLUGGED INTO A UL APPROVED THREE
CONTACT ELECTRICAL OUTLET; AND ALL AUXILIARY
COMPONENTS (EXCLUDING MONITORS) MUST BE PLUGGED INTO
THE ISOLATION TRANSFORMER UNIT SUPPLIED WITH THE
SYSTEM. (PART #735–045–G1 FOR 115V, PART #735–045–G2 FOR
230V, PART #735–045–G3 FOR 100V). DO NOT SIMPLY PLUG THE
GOLDONE SYSTEM INTO A WALL PLUG!
DO NOT
POSITION OR USE THIS SYSTEM IN A LOCATION WHERE IT
MAY BE SUBJECT TO SPLASHED OR SPILLED LIQUIDS!
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).