om 5184516-100 Rev. 5 D-4
2-4 How to recognize when this mode is selected
• The “Dig Pulsed” button on the IUI interface screen will be selected when this
mode is engaged.
• Any mode that uses the Flouroscopy tube (tube 2) will display the fluoro “eye”
icon on both the IUI and IDD when selected/engaged:
2-5 Example of specific clinical procedures
Pulsed fluoro is used as a dose reduction feature and is helpful in longer studies
in which dose savings is desired, as well as pediatric applications.
2-6 How mode should be used
Used by or under the direction of a physician specializing in diagnostic imaging.
SECTION 3
DIGITAL SUBTRACTION ANGIOGRAPHY (DSA) MODE
3-1 Description
This mode is a purchasable option.
Digital Subtraction Angiography (DSA) is the procedure to visualize blood vessels
with contrast medium in a bony environment by subtracting the pre-contrast
image (the mask) from the image with contrast medium.
3-2 System Controlled Technique Factors
• Fixed or Automatically Selected Technique Factors: Allows for the subtraction
of images frames resulting in the frames difference being displayed.
• How automatic adjustment is controlled: A calibrated ion chamber controls the
duration of the exposure time for each frame.
3-3 How to engage/ disengage
This mode is engaged by selecting a DSA protocol on the IUI interface screen.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).