om 5184516-100 Rev. 5 D-4
2-4 How to recognize when this mode is selected
The “Dig Pulsed” button on the IUI interface screen will be selected when this
mode is engaged.
Any mode that uses the Flouroscopy tube (tube 2) will display the fluoro “eye”
icon on both the IUI and IDD when selected/engaged:
2-5 Example of specific clinical procedures
Pulsed fluoro is used as a dose reduction feature and is helpful in longer studies
in which dose savings is desired, as well as pediatric applications.
2-6 How mode should be used
Used by or under the direction of a physician specializing in diagnostic imaging.
SECTION 3
DIGITAL SUBTRACTION ANGIOGRAPHY (DSA) MODE
3-1 Description
This mode is a purchasable option.
Digital Subtraction Angiography (DSA) is the procedure to visualize blood vessels
with contrast medium in a bony environment by subtracting the pre-contrast
image (the mask) from the image with contrast medium.
3-2 System Controlled Technique Factors
Fixed or Automatically Selected Technique Factors: Allows for the subtraction
of images frames resulting in the frames difference being displayed.
How automatic adjustment is controlled: A calibrated ion chamber controls the
duration of the exposure time for each frame.
3-3 How to engage/ disengage
This mode is engaged by selecting a DSA protocol on the IUI interface screen.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Precision 500D and is the answer not in the manual?

GE Precision 500D Specifications

General IconGeneral
BrandGE
ModelPrecision 500D
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals