om 5184516-100 Rev. 5 5-27
The protocol database file is stored in the same location on the system as the
other system log files.
ILLUSTRATION 5-26
CHANGES TO PROTOCOLS
3-2-1 Parameters for Tube 1 (Fluoro) Protocol Step
For the Tube 1 Protocol Step, there is only one Receptor and there is no Patient
Size. An initial EPT (Estimated Patient Thickness) is used, with the contrast media
selected. The initial EPT is used for the first exposure and then subsequent expo-
sures use the Patient Size calculated by exposure management. The initial or
default EPT is hard coded in system software.
For the Tube 1 Protocol Step, the technique parameters that is defined by the
user for digital record are: Contrast Media, AEC / Angio, and Record Frame Rate.
If Angio, then Sub / No Sub is also defined. The technique parameters that is
When loading the default protocols the user must select the correct
Protocol CD-Rom.
The system will warn the user that specific protocols have had
modifications made to them. The user must be aware of any user-
defined protocols before accepting installation of any
Protocol CD-
Roms during Protocol Retrieve. Refer to ILLUSTRATION 5-26.
If protocols have been modified, and a “Default” CD is used to restore
system configuration, all
of the user-defined protocols will be deleted
and must be redefined.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).