Rosemount Model 1195/ProPlate/Mass ProPlate
7-4
Figure 7-2.
ProPlate Electronics with
Optional LCD Meter.
Diagnostic Messages In addition to the output, the LCD meter displays abbreviated
operation, error, and warning messages for troubleshooting the
flowmeter. Messages appear according to their priority, with normal
operating messages appearing last. To determine the cause of a
message, use a Model 275 HART Communicator to further interrogate
the flowmeter. A description of each LCD diagnostic message follows.
Error
Error messages appear on the LCD meter display to inform you of
serious problems effecting the operation of the flowmeter. The meter
displays an error message until the error condition is corrected, and the
analog output is driven to the specified alarm level. No other flowmeter
information is displayed during an alarm condition.
FAIL
The flowmeter CPU board and the sensor module are incompatible.
If you encounter this message, contact Dieterich Standard Inc. at
303-530-9600 if you need assistance.
FAIL MODULE
The sensor module is disconnected or is malfunctioning. Verify that the
sensor module ribbon cable is connected to the back of the electronics
board. If the ribbon cable is properly connected, there is a problem
within the sensor module. Possible sources of problems include:
• Pressure or temperature updates are not being received in the
sensor module
• A non-volatile memory fault that will effect flowmeter operation
has been detected in the module by the memory verification
routine
Some non-volatile memory faults are user-repairable. Use a Model 275
HART Communicator to diagnose the error and determine if it is
repairable. Any error message that ends in “FACTORY” is not
repairable. In cases of non user-repairable errors, you must replace the
sensor module. See Remove the Flowmeter from Service on page 9-2, or
contact Dieterich Standard Inc. at 303-530-9600 if you need assistance.