Rosemount Model 1195/ProPlate/Mass ProPlate
10-12
Performance
Specifications
Differential Pressure Ambient Temperature
Effect Per 50°F (28°C)
0.025% of URL + 0.175% of span.
Spans from 1:1 to 30:1.
0.035% of URL – 0.125% of span.
Spans from 30:1 to 100:1.
Static Pressure Effects
Zero error =
0.10% of URL per 1,000 psi (6894 kPa).
Span error =
0.20% of reading per 1,000 psi (6894 kPa).
Stability
0.1% of URL for 12 months.
Absolute/Gage Pressure Ambient Temperature
Effect Per 50°F(28°C)
0.05% of URL + 0.175% of span.
Spans from 1:1 to 30:1.
0.06% of URL-0.125% of span.
Spans from 30:1 to 100:1.
Stability
0.1% of URL for 12 months.
Process Temperature Ambient Temperature
Effect Per 50°F (28°C)
0.36°F (0.20°C) for process temperatures from –40 °F to 185°F (–
40°Cto 85°C).
(0.64°F (0.36°C) + 0.16% of reading) for process temperatures from
185°F (85°C) to 400°F (204°C).
Physical Specifications Electrical Considerations
½–14 NPT, PG 13.5, and CM20 conduit. HART interface connections
permanently fixed to terminal block.
Process-Wetted Parts
Sensor
316L SST.
Integral Manifolds
316 SST.
Remote Manifolds
316 SST or CS.
Electronics Vent Valves and Process Flanges
316 SST.