EasyManuals Logo

Emerson Rosemount 1195 User Manual

Emerson Rosemount 1195
229 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #175 background imageLoading...
Page #175 background image
Rosemount Model 1195/ProPlate/Mass ProPlate
10-12
Performance
Specifications
Differential Pressure Ambient Temperature
Effect Per 50°F (28°C)
0.025% of URL + 0.175% of span.
Spans from 1:1 to 30:1.
0.035% of URL 0.125% of span.
Spans from 30:1 to 100:1.
Static Pressure Effects
Zero error =
0.10% of URL per 1,000 psi (6894 kPa).
Span error =
0.20% of reading per 1,000 psi (6894 kPa).
Stability
0.1% of URL for 12 months.
Absolute/Gage Pressure Ambient Temperature
Effect Per 50°F(28°C)
0.05% of URL + 0.175% of span.
Spans from 1:1 to 30:1.
0.06% of URL-0.125% of span.
Spans from 30:1 to 100:1.
Stability
0.1% of URL for 12 months.
Process Temperature Ambient Temperature
Effect Per 50°F (28°C)
0.36°F (0.20°C) for process temperatures from 40 °F to 185°F (
40°Cto 85°C).
(0.64°F (0.36°C) + 0.16% of reading) for process temperatures from
185°F (85°C) to 400°F (204°C).
Physical Specifications Electrical Considerations
½–14 NPT, PG 13.5, and CM20 conduit. HART interface connections
permanently fixed to terminal block.
Process-Wetted Parts
Sensor
316L SST.
Integral Manifolds
316 SST.
Remote Manifolds
316 SST or CS.
Electronics Vent Valves and Process Flanges
316 SST.

Table of Contents

Other manuals for Emerson Rosemount 1195

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount 1195 and is the answer not in the manual?

Emerson Rosemount 1195 Specifications

General IconGeneral
BrandEmerson
ModelRosemount 1195
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals